Medical Education Systems, Inc. is one of the most cost effective CEU solutions online, and is backed by a strong history of innovation, customer service and easy-to-earn distance learning courses.Each nursing CEU course is offered online or via the postal service, and you are welcome to take the test as many times as necessary to pass the course with no additional fees.
Wednesday, May 11, 2011
Adapt SV and Adaptive Servo-Ventilation
VPAP® Adapt SV and Adaptive Servo-Ventilation
With its advanced Adaptive-Servo Ventilation (ASV) algorithm, the ResMed VPAP Adapt SV provides ventilatory support to rapidly treat all forms of central sleep apnea (CSA), mixed apnea
and periodic breathing, commonly known as Cheyne-Stokes respiration (CSR).
The ASV algorithm delivers customized therapy
Ventilation to a moving target
To determine the degree of pressure support needed, the ASV algorithm continuously calculates a target ventilation. Based on respiratory rate and tidal volume, the target is 90% of the patient’s recent average ventilation—that means that ventilation can vary gradually and naturally over the course of the night.
Patient/machine synchronization
The algorithm uses three factors to achieve synchronization between pressure support and the patient’s breathing.
1. The patient’s own recent average respiratory rate—including the ratio of inspiration to expiration and the length of any expiratory pause.
2. The instantaneous direction, magnitude, and rate of change of the patient’s airflow, which are measured at a series of set points during each breath.
3. A backup respiratory rate of 15 breaths per minute.
To ensure ventilatory support is synchronized to the patient’s effort, the VPAP Adapt SV relies on factors one and two. When a central apnea/hypopnea occurs, support initially continues to reflect the
patient’s recent breathing pattern. However, as the apnea/hypopnea persists, the device increasingly uses the backup respiratory rate.
Minimal support during stable breathing
The ASV algorithm starts working even before central sleep apnea events occur, from the moment the patient lies down, puts on the mask and switches on the VPAP Adapt SV.
So long as ventilation is at or above the target:
• the magnitude of the pressure support remains minimal—the default minimum value of 3 cm H20 provides enough support to do about half the work of breathing. • the underlying end expiration pressure (EEP) is clinician adjustable from 5 to 10 cm H20, helps reduce obstructive events and can also reduce central events.
Support when it’s needed
When a central hypopnea/apnea occurs and ventilation suddenly drops below the target:
• pressure support rapidly increases over a few breaths to
keep ventilation at the target
• the default maximum pressure support value of 10 cm H20 will fully ventilate a centrally apneic patient with an open airway and normal lungs.
Adaptive support when breathing resumes
As breathing resumes and total ventilation exceeds the target:
• pressure support is rapidly reduced back towards the minimum 3 cm H20
• this reduces the likelihood of over-ventilation and hypocapnia,
Why ASV instead of bilevel?
Most CSA/CSR patients have trouble tolerating conventional bilevel ventilatory support. The pressure must be adjusted to a constant high level to adequately support a patient during apnea/hypopnea.
This leads to over-ventilation during periods of normal breathing or hyperpnea, which causes arousals and discomfort. It can even propagate more CSA events.
The ASV algorithm stabilizes patient breathing
By ventilating the patient appropriately during periods of hypopnea and apnea and reducing support during periods of hyperventilation and normal breathing, the ASV algorithm rapidly stabilizes breathing patterns and arterial blood gases and minimizes discomfort and arousals often associated with bilevel treatment. which can lead to vocal cord closure and further apneas. http://www.mededsys.com/product_info.php?cPath=1&products_id=201
Subscribe to: